 Model 1010 TOC Analyzer  General specifications  Performance Specifications
Communications Requirements Options
 Model 1010 TOC Analyzer 

The Model 1010 Total Organic Carbon [TOC], Analyzer is an automated system for analyzing aqueous samples for total inorganic carbon [TIC] and TOC. The Model 1010 performs both TIC and TOC analysis on one aliquot of sample. This allows TOC to be measured directly rather than through subtraction [TC-IC], as is the case with most TOC analyzers.

The Model 1010 uses approved 100℃ per sulfate oxidation  methodology for analyzing samples containing 2ppb to 10,000ppm of organic carbon. This heated per sulfate methodology has been proven to be the most reliable technique for acidic, ionic, refractory, and other difficult matrices. In addition, 100℃ per sulfate oxidation provides recoveries better than or equal to UV-promoted oxidation without the cost and maintenance of UV lamp replacement.

Total Organic Carbon Analyzer

Measures total carbon [TC], total inorganic carbon [TIC], and total organic carbon [TOC].

  • Same-sample TIC/TOC measurement.
  • Small, compact size.
  • Ultralow-level sensitivity.
  • Single-or multipoint calibration
  • sequence programming for unattended operation.
  • Programmable rinse capability.
  • Nondispersive infrared [CO2 specific] detector.
  • For laboratory or at-line applications.
  • Choice of interface options: keyboard and monitor, or Windows00-based software [WinTOCTM], for data han-dling and analyzer control.

Principle of Operation: Same-sample determination of TIC and TOC is performed by wet oxidation. TIC is determined by acidifying the sample to convert inorganics to carbon dioxide. The Carbon dioxide is purged from the sample and then detected by a nondispersive infrared detector [NDIR] calibrated to directly display the mass of carbon dioxide measured. This mass is proportional to the mass of TIC in the sample.

After TIC is removed from the sample, per sulfate reagent is added. The per sulfate reacts with organic carbon in the sample at 100℃ to form carbon dioxide, which is purged from the sample and detected by an NDIR. The resulting mass of carbon dioxide is proportional to the mass of TOC in the sample.

Principal Applications:
  • Standard Method 5310C                           Boiler feedwater
  • USEPA 4151                                      Cooling water
  • USEPA 9060                                      Seawater
  • ASTM D4839                                      Drinking water
  • ASTM D4779                                      Wastewater
  • Cleaning validation                             Ultrapure water
  • USP 643[WFI]                                    Groundwater

return     


General specifications 

Dimensions 

  • With keyboard and monitor:24.75″H×13.5″W×19″D(62.9cmH×34.3cmW×48.3cmD)
  • With WinTOC: 17″H×13.5″W×19″D(43.2cmH×34.3cmW×48.3cmD)
  • With autosampler option: 34″H×18.75″W×23.5″D(86.4cmH×47.6cmW×59.7cmD)

Weight

  • 46lbs(20kg)
  • Autosampler option:22 lbs (10kg)

return     

Performance Specifications

Method

  • TOC:100℃ per sulfate (Na2S2O8) oxidation (TOC measured directly after removal of TIC)
  • TIC: Acidification (H3PO4) and sparging

Analysis Modes

  • TIC,TOC (NPOC), TC, and TIC/TOC
  • Purgeable organic carbon (POC): available as part of TC measuerment 

Sample Introduction 

  • Syringe (manual)
  • Loop sampling
  • On-line/at-line
  • Vial autosampler

Sample Size

  • By syringe: 1 μl to 25 mL
  • By loop: 1-10 mL in 1- mL increments, 5-25 mL in 5-mL increments
  • Allowable particulate size: 200 μm

Range

Loop Sampling (sipper/on-line/autosampler)

  • TOC:2ppbC to 125ppmC
  • TIC:2ppbC to 125 ppmC

Syringe Injection (manual)

  • TOC: up to 100,000ppmC
  • TIC: up to 100,000ppmC

Detection

  • Nondispersive infrared (NDIR)

Precision

  • TOC: Greater of 2% RSD or ±2ppbC
  • TIC: Greater of 2% RSD or ±2ppbC

Time of Analysis

  • TIC: 3.5 min
  • TOC:4 min
  • TC: 4 min
  • TIC/TOC: 7-9 min (typical)

return     

Communications

  • Parallel and serial communications (RS-232-C)
  • Auxiliary output for optional equipment

Data Storage

  • Keyboard and monitor: no values stored for recall (printer output only)
  • WinTOC software: data file storage, processing, and recall (limited only by drive space)

Data Export Format

  • ASCII format
  • Selectable delimiters
  • Easy import into most LIMS systems

Requirements

Gas Requirements

  • Nitrogen,99.98% purity or better, 50-60psig (345-4132kPa)
  • Consumption: 300mL/min

Reagent Requirements

  • Sodium per sulfate (Na2S2O8), 100-200g/L (uses 1mL/sample)
  • Phosphoric acid (H3PO4), 5%(uses 0.2mL/sample)

Power Requirements

  • 100(±10%) VAC 50/60 Hz
  • 110-125(±10%) VAC 50/60 Hz
  • 210-240(±10%) VAC 50/60 Hz (optional)

Ambient Temperature

  • 0°-40℃

return     

Options

User Interface Options

  • Keyboard and monitor(9-inch)requires printer for data output
  • Win TOc (Windows-based) software for PC control (computer not included)

Other Options

  • Printer (dot-matrix)
  • Vial autosampler
  • Halide scrubber option
  • Solids module

return     

   
   
   